LINE REGULATOR
Número de grupo: 3611Whole medical range is available on
A stabilizer is a pressure reduction unit with the task to equalize the eventual pressure variation in the hospital pipeline system to ensure a correct pressure from the terminal units.
The stabilizer makes it possible to distribute gas with different pressure to department and buildings in the hospital area. In some cases it is needed to deliver a higher pressure from the main gas manifold to compensate for small pipe dimensions. In those cases the Stabilizer shall be mounted as close as possible before the first terminal unit, to ensure a correct pressure to the patient.
FOR A COMPLETE LINE REGULATOR ADD:
- Plastic cover for locking
- Alarm unit (included if ordered together with HP unit)
Item No. | Denomination | Gas type |
Inlet* |
0727333 | LINE REG | O2, N2O, Air, CO2, N2 | LH |
K141621 | LINE REG | O2 – AFNOR | LH |
K141631 | LINE REG | O2 – AFNOR | RH |
K141622 | LINE REG | N2O – AFNOR | LH |
K141632 | LINE REG | N2O – AFNOR | RH |
K141623 | LINE REG | Air – AFNOR | LH |
K141633 | LINE REG | Air – AFNOR | RH |
K141629 | LINE REG | Air-800 – AFNOR | LH |
K141639 | LINE REG | Air-800 – AFNOR | RH |
K141624 | LINE REG | N2 – AFNOR | LH |
K141625 | LINE REG | CO2 – AFNOR | LH |
* LH = inlet from left side; RH = inlet from right side
735100001553 Line Regulator leaflet EN
TECHNICAL DATA
Gases: | O2, Air, Air-800, N2, N2O, CO2, (all medical gases) |
Nominal flow: | 40 m3/h |
Inlet nominal pressure: | 16 bar (1600 kPa) |
Outlet nominal pressure: | 4,5 bar (setting range 0,5–10 bar) |
Inlet connection: | G1/2"M+ soldering piece pipe ∅ 12 mm |
Outlet connection: | G1/2"M+ soldering piece pipe ∅ 12 mm |
Pressure sensors: | Optional (Pressure switches; Transmitters 0–50 mV; 4–20 mA |
Emergency QC inlet: | Optional QC by national standards |
Regulatory status:
|
Complies with Medical Devices Directive 93/42/EEC present HTM 02-01 |